<cite id="dfb1x"></cite>
<span id="dfb1x"></span>

                  Article Position:Home > article > Documentation > Detailed
                  Kinematic Viscosity Meter
                  Date:2016-01-18   Hit:976

                  運動粘度測定儀

                    Kinematic Viscosity Meter

                  ModelND-3S

                  使用說明書V3.0Instruction

                   

                   

                  大慶市日上儀器制造有限公司

                  Daqing City Instrument Manufacture Co., Ltd.

                  公司地址:大慶高新區產業三區禾豐路東段

                  Address:East Part of Hefeng Road, No.3 Industrial District,High-tech Development Area, Daqing

                  電話:0459-8186000     TEL:0459-8186000

                  傳真:0459-8186000     FAX:0459-8186000

                  網址websitehttp://www. rshang.com/

                      Content

                  目錄Preface. 1

                    Preface. 2

                  第一部分 概述General Introductions 4

                  1.術語Terminology 4

                  2.技術指標 Technical Index 4

                  2.1技術參數Parameters 4

                  2.2 儀器特點Features 5

                  3. 結構 Structure 6

                  3.1 運動粘度測定儀主視圖Front View.. 6

                  3.2 運動粘度測定儀后面板視圖Back View.. 7

                  4 安裝Installation 7

                  4.1 開箱Unpack. 7

                  4.2 儀器安裝Installation. 8

                  第二部分 ND-3S操作說明Operating Instructions 9

                  1 鍵盤功能Keyboard Function 9

                  2 開機Start Up 10

                  3 菜單Menu 12

                  4 試驗步驟Procedures 16

                  5操作注意事項Attentions 17

                  第三部分 儀器維護、保養及安全說明Maintenance&Safety Instructions 18

                  1儀器維護與保養Maintenance 18

                  2粘度計管清洗cleaning viscometer 18

                  3安全說明Safety Instructions 18

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                    Preface

                  ND系列運動粘度測定儀,是大慶市日上儀器制造有限公司自主開發的具有國內領先水平的高科技產品。本儀器是依據標準ASTM D445-06設計制造的專用測儀試器,適用于測定液體石油產品(指牛頓液體)的運動粘度。

                  ND series kinematic viscosity tester is a high-tech product that ranks the top in domestic and independent developed by Daqing City Rishang Instrument Manufacture Company Limited. It is specific tester that design and produced according to ASTM D445-06and used for determination of kinematic viscosity of liquid petroleum products(refers to Newtonian liquids).

                  本儀器具有計時試樣運動時間,自動計算運動粘度的最終結果。本方法適用于測定液體石油產品(指牛頓液體)的運動粘度,其單位為m2/s,通常在實際中使用為mm2/s。動力粘度可由測得的運動粘度乘以液體的密度求得。本方法是在某一恒定的溫度下,測定一定體積的液體在重力下流過一個標定好的玻璃毛細管粘度計的時間,粘度計的毛細管常數與流動時間的乘積,即為該溫度下測定液體的運動粘度。該溫度下運動粘度和同溫度下液體的密度之積為該溫度下的動力粘度。

                  The apparatus have the timing function of samples excises, and automatically calculate the final results of kinematic viscosity. This method fits for testing kinematic viscosity of liquid petroleum products (refers to Newtonian liquids), its unit is m2/s, practically used as mm2/s. dynamic viscosity is the product of measured viscosity and liquid density.

                  Under a given hydrostatic temperature, test the time which a fixed volume of fluid flow through glass capillary viscometer, the product of capillary coefficient and flow time namely the liquid kinematic viscosity under the temperature. The product of kinematic viscosity and liquid density is the dynamic viscosity under the temperature.

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                  第一部分 概述

                  Chapter One    General Introductions

                  1.術語Terminology

                       粘度:液體受外力作用移動時,液體分子間產生內摩擦力的性質,稱為粘度。

                       ViscosityFlow of a fluid under external forceinner friction between molecular called viscosity.

                  運動粘度:指在某一恒定溫度下,測定一定體積的液體在重力下流過一個標定好的玻璃毛細管粘度計的時間,粘度計的毛細管常數與流動時間的乘積,即為該溫度下測定液體的運動粘度。在溫度t時運動粘度用符號Vt表示。

                       Kinematic viscosity: Under a given hydrostatic temperature, test the time which a fixed volume of fluid flow through glass capillary viscometer, the product of capillary coefficient and flow time namely the liquid kinematic viscosity under the temperature. “Vt” stands for viscosity at“t”temperature.

                       動力粘度:某溫度下運動粘度和同溫度下液體的密度之積為該溫度下液體的動力粘度。在溫度t時的動力粘度用符號ηt表示。

                       Dynamic viscosity: under the same temperature, the product of kinematic viscosity and liquid density is the dynamic viscosity, it is indicates by the symbol “ηt”.

                  意義:粘度是絕緣油的重要指標之一,使用于變壓器的絕緣油,在不影響油品其他指標(如閃點)的前提下,粘度盡可能的低一些為好,因為粘度越低,變壓器的循環冷卻效果越好;同時粘度也是潤滑油的最重要指標之一,因為潤滑油的牌號大部分是以油品標準中的運動粘度的平均值來區分的,另外正確的選擇一定粘度的潤滑油,可保證發電機和汽輪機組穩定可靠的運行狀態,粘度過大或過小都會改變發動機的功率,增大燃料的消耗,增加機器磨損等;

                       Significance: viscosity is one of important index of insulating oil, the viscosity of the transformer insulating oil should as lower as possible, because the lower viscosity, the better circulating cooling effects of transformer; meanwhile, it is one of important index of lubricant oil, the number of lubricant oil generally distinguished by the average of kinematic viscosityin addition, proper choose of viscosity of lubricant oil will guarantee stable running of generator and turbine sets, viscosity is too higher or lower will change the power of generator, increase fuel consumption and wear etc.

                  2.技術指標 Technical Index

                  2.1技術參數Parameters

                  控溫范圍:室溫~100

                  Temperature control range: room temperature~100

                  ☆ 控溫精度±0.1

                  precision±0.1

                  測試單元:4

                  Testing unit4 channels

                  存儲數據:100/

                  Data storage100 pcs

                  環境溫度 540(室內)

                  Ambient temperature5~40indoor

                  存放溫度 :-1055(室內)

                  storage temperature-10~55℃(indoor

                  相對濕度:10~80%

                  relative humidity10~80%

                  220VAC ±10% 50Hz

                  power supply220VAC±10%50Hz

                    消耗功率 <1000W

                    power consumption:<1000W

                    外型尺寸:570×370×500mm

                    size570×370×575mm

                  2.2 儀器特點Features

                  符合標準 GB/T265-88

                  According to standard: ASTM D445-06

                  適用于測定液體石油產品的運動粘度。

                  Apply to test of kinematic viscosity of liquid petroleum products

                  液晶圖文顯示。

                  Liquid crystal screen displays of pictures and words.

                  單片微機自動控制溫度。

                  Automatically control temperature by microcontroller.

                    不掉電時鐘,手動啟停計時。

                    Uninterruptible power clock, manually start/stop timing.

                    自動運算運動粘度值,數據存儲及打印。

                    Automatically calculate viscosity value, data storage and print.

                  3. 結構 Structure

                  3.1 運動粘度測定儀主視圖Front View

                  3.2 運動粘度測定儀后面板視圖Back View

                  picture1-2

                  1、電源開關switch 2、電源插座supply socket  3、保險  fuse

                  注意:系統組成部分的插圖可能與實物有細微差別,以實物為準。

                  Attention: there is subtle difference between the picture and the object, please take the real object as standard.

                  4 安裝Installation

                  4.1 開箱Unpack

                  標準附件:Standard attachments:

                  粘度管架  capillary viscometer rack            4set

                  防塵蓋  dust-proof cover                     4pcs

                  保險管10A    fuse pipe                     2pcs

                  電源線     supply line                      1 pcs

                  打開包裝后,最好使其在實驗室放置2小時以上方可通電,(特別是在低溫存儲的情況下)

                  open the packageyou’d better place the apparatus in laboratory over two hours and electrified.

                  ◇ 使其放置在光線較好的水平平臺上,以便觀察試樣的流動界面

                  Put it on the platform with bright light in order to observe flow of fluid.

                  ◇ 儀器的背面和側面隔墻的最小間隙為200mm

                  The minimum partition between the back view and the side view is 200mm.

                  ◇ 要防止水飛濺到儀器或其他細小雜物掉到儀器上

                  Prevent the water from splashing or other tiny materials drop on the apparatus.

                  4.2 儀器安裝Installation

                  把儀器擺放在工作平臺上,根據測定溫度的要求在浴缸內注入相應的浴液體至水位標記處。

                  Place the apparatus on the working platform, according to test temperature requirement add corresponding bath liquid into the bath vat until reach the marked level.

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                  第二部分 ND-3S操作說明

                  Chapter Two ND-3S Operating Instructions

                  警告:Warning:

                    操作者必須閱讀本章內容,并按要求進行操作,以免操作不當造成儀器部件的損壞

                    Operators must read this chapter and worked as requiredso as to avoid break the parts due to improper use.

                    非操作人員必須得到許可,方能操作本儀器

                     Non-operators have to get permission to operate the apparatus.

                  該儀器采用液晶顯示屏,界面友好,直觀明了,用戶在使用過程中,可用鍵盤進行操控,方便快捷。

                     Our product adopts touch liquid crystal displayfriendlyintuitiveduring the process of usecustomers are capable of operate the apparatus by keyboard, convenient and rapid.

                  1 鍵盤功能Keyboard Function

                  19 [數字]鍵,在設定時用于數據輸入,其中14鍵在允許計時情況下用于14號粘度

                  計的試樣流動時間的計時開始和結束。

                  while setting,1~9numberkey used for input data; On condition that timing permits, 1~4 key used for 1~4 viscometer start and stop timing of flow time.

                  [小數點]鍵,用于數據輸入。

                  dotkey used for input data

                  [退格]鍵,用于數據刪除,每按一次,刪除一位數據。

                  back key used for delete dataevery click will delete one bit of data

                  [上]鍵,向上選擇,用于菜單選擇,查詢數據。

                  upkeyupward selectused for menu select and query data

                  [下]鍵,向下選擇,用于菜單選擇,查詢數據。

                  downkey ,downward selectused for menu select and query data

                  [左]鍵,左向選擇,用于菜單選擇。

                  leftkey, leftward selectused for menu select

                  [右]鍵,右向選擇,用于菜單選擇。

                  rightkey, rightward selectused for menu select

                  [打印]鍵,打印測試數據和歷史數據。

                  printkeyprint testing data and historical data.

                  [計時]鍵,在恒溫情況下按此鍵允許計時。

                  timekey used for timing under constant temperature

                  [結束]鍵,結束試驗,清除實驗數據。

                  endkey used for end experiment, clear the data .

                  [確定]鍵,用于數據輸入結束或執行所選菜單命令。

                  enterkey used for finish the input or conduct the selected order

                  [菜單]鍵,用于進入功能菜單,按一次進入,再按一次退出菜單。

                  menukey used for enter or quit the menu

                  [數據]鍵,用于查詢以往測試數據。

                  datakey used for query of past testing data

                  [幫助]鍵,用于查詢儀器相關信息。

                  helpkey used for query of relevant information about product

                  [復位]鍵,用于儀器程序重新啟動,一般用于儀器工作異常時。

                  resetkey used for restart the programusually when the apparatus behaves abnormal.

                  2 開機Start Up

                  打開電源開關,液晶顯示屏上顯示開機界面。如下:

                  Turn on the powerLCD shows of startup interfaceas the following

                  picture2-1

                      在開機界面中,可查閱儀器型號、儀器名稱、軟件版本、公司名稱、公司網址。

                  In startup interface, you can search for instrument modelnameversioncompany name and website.

                  開機界面持續幾秒鐘后,儀器進行自檢,啟動試驗程序,進入主界面。如下:

                  Hold several seconds after startup, the apparatus conducts self-check and launch test program, then enter into main menu. As the following:

                  NoteNo stands for Forbid timing

                  picture2-2

                  在上述圖中,SV表示設定溫度,PV表示實測溫度,加熱中……表示儀器現在的工作狀態,禁止計時表示此時不能進行計時操作。如果設定溫度與所做油品所需要的溫度不相符時,可用鍵盤對菜單進行操作,更改設定溫度。

                  In above picture, SV stands for set temperature, and PV indicates measure temperature, heating shows the current working state of apparatus, forbid timing means can’t time at this time. when the setting temperature isn’t consistent with the required temperature, you can operate the menu though keyboard, change the setting temperature.

                  當實測溫度到達設定溫度后,儀器自動轉入檢測界面。如下:

                  When the measured temperature equals to the setting temperature, the apparatus automatically link to the next interface. It shown in the following:

                  NoteCV1stands for capillary viscometer 1

                  picture2-3

                  在此界面下按計時鍵,進入允許計時界面,如2-4

                  In this interface, click [time]key will enter into permit timing interface, as shown in picture2-4.

                  picture2-4

                  剛跳入此界面時,光標停在“管一”處,此時按][鍵可選擇毛細管,按“確定”鍵,進入相應毛細管1試驗界面。如下:

                  The cursor stop at “CV1” when linked to this interface, at this time, press[left][right]key can choose capillary, and clickenter” key will link to corresponding capillary test interface. As the following:

                  picture2-5

                   在此界面下,如做好所有準備工作,當試樣液面回落到上刻度時按下“1鍵開始計時,當試樣液面回落到下刻度時再次按下“1鍵停止計時并鎖定時間值。如此反復測試四組試樣的流動時間后,自動計算出其平均流動時間和粘度值。測試結束后按下“打印”鍵可打印當前顯示的測試數據。

                  Under this interface, if ready for all, click1start to timing when the liquid level fall back to the upper scale, click1again to stop timing and locked the time when the liquid level fall back to the lower scale. Just like this repeatedly test the flow time of 4 teams samples, automatically calculate the average flow time and viscosity value. After finishing the experiment, clickprintwill print the current displayed data.

                  在此過程中可穿插進行其它毛細管的試驗,操作方法相同,按對應的數字鍵進入相應毛細管界面并控制起停,按][鍵可切換顯示不同毛細管的試驗數據。

                  During this process, you can alternately test other capillary with the same method, press corresponding number key will enter into relevant capillary interface and control the start/stopclick[left][right] can switch to different displays of different capillaries.

                  3 菜單Menu

                  按下“菜單”鍵進入菜單界面。如下:

                  Pressmenuwill link to menu interface. As the following:

                  picture2-5

                  3.1設置setup

                      進入菜單界面光標停在“參數設定”處,此時按“確定”鍵進入設定界面。用于設置實

                  驗參數,根據不同的樣品及粘度計管設置所需參數,對程序進行控制。如下:

                  After entered menu interfacethe cursor stop at setup”,now click enterwill link to setting interface. It used for set required parameters according to different samplesviscometers, and control the program. As the following:

                  picture2-6

                  3.1.1進入菜單界面光標停在“設定溫度”處,此時按“確定”鍵進入設定溫度界面。如下:

                  After entered into the interface and the cursor stop at “temperature”, now click “enter” will link to setting temperature interface. As the following:

                  System temperature setting

                  Setting temperature:

                  Automatic

                  picture2-7

                  在設定溫度界面下通過鍵盤數字按鍵對溫度進行設定,設定后必須按“確定”鍵對設定值進行記憶。

                  Under this interface, you can set the temperature through press key of number, after setting, you must press “enter” for save all settings.

                   

                  3.1.2在菜單界面下用移位按鍵將光標移到 “毛細管常數”處,此時按“確定”鍵進入毛細管常數設定界面。如下:

                  After entered into the interface and the cursor stop at “capillary coefficient”, now click “enter” will link to capillary coefficient setting interface. As the following:

                  CV1-C stands for CV1 coefficient

                  picture2-8

                  在毛細管常數設定界面下通過鍵盤數字按鍵對毛細管常數進行設定,設定后必須按“確定”鍵對設定值進行記憶。默認設定為毛細管1常數,可通過左右移位按鍵切換到其它毛細管常數進行設定。

                  Under this interface, you can set the coefficient through press key of number, after setting, you must press “enter” for save all settings. The default setting is capillary 1, and switch to other capillary setting through [left] [right] key.

                  3.1.3在菜單界面下用移位按鍵將光標移到 “系統時間”處,此時按“確定”鍵進入系統時間設定界面。如下:

                  After entered into the interface and the cursor stop at “system time”, now click “enter” will link to system time setting interface. As the following:

                  picture2-9

                  系統時間設定界面下通過鍵盤數字及左右移位按鍵對系統時間進行設定,設定后必須按“確定”鍵對設定值進行記憶。

                  Under this interface, you can set the time through press key of number, after setting, you must press “enter” for save all settings.

                   

                  3.1.4在菜單界面下用移位按鍵將光標移到 “溫度微調”處,此時按“確定”鍵進入溫度微調設定界面。如下:

                  After entered into the interface and the cursor stop at “T correction”, now click “enter” will link to temperature fine tuning setting interface. As the following:

                  picture2-10

                  此功能是對溫度測量出現誤差時的數據平移修正

                  When the temperature appears deviation just use this function to amend data.

                  溫度微調設定界面下通過鍵盤數字按鍵對溫度微調進行設定,設定后必須按“確定”鍵對設定值進行記憶。控制溫度穩定后,屏幕顯示的溫度與實際溫度有誤差時,在此調節。首先移動光標到微調數據處,如果實際溫度低于屏幕顯示的溫度時應進行正微調,按上下移位按鍵將微調值前綴符號調整為“+”號,微調值通過數字按鍵進行設定;如果實際溫度高于屏幕顯示的溫度應進行負微調,按上下移位按鍵將微調值前綴符號調整為“—”號,微調值通過數字按鍵進行設定。當微調值到達期望值后,按“確認”鍵進行數據記憶,微調完畢。

                  Under this interface, you can fine tuning the temperature through press key of number, after setting, you must press “enter” for save all settings. After controlled temperature well stabled, there still have bias between displayed temperature and practical temperature, just tune it here. Firstly, move the cursor to fine tuning data, if the practical temperature lower than displayed, make positive fine tuning, change the prefix symbol into “+” by shifting “Ù” “Ú” key, fine tuning value set by number key; if the practical temperature higher than displayed, make positive fine tuning, change the prefix symbol into “-” by pressing shift key of“Ù” “Ú”, fine tuning value set by number key. When the fine tuning value reached the expectation, click “enter” to save all settings, fine tuning is completed.

                  3.2數據查詢data query

                  在菜單界面下通過移位按鍵將光標移到“數據查詢”處,此時按“確定”鍵進入數據查詢界面。在數據查詢界面中,可查閱歷史數據,每路可存儲100組數據。如下:

                  After entered into the interface and the cursor stop at “data query”, now click “enter” will link to inquiry interface. Under this interface, you can look up historical data, and every channel can save 100 pieces of data. As the following:

                  picture2-11

                  數據查詢界面下通過移位按鍵對毛細管1~4數據進行查詢,進入數據查詢界面光標停在“毛細管1數據”處,此時按“確定”鍵進入毛細管1數據查詢界面。如下:

                  Under data query interface, you can inquiry of data from1 to 4 by shifting key, entered into the interface and the cursor stop at “CV1 data”, now clickenterwill link to data query  interface. As the following:

                  picture2-12

                  毛細管1數據查詢界面中顯示毛細管1的所有測試數據。

                  In CV1 data query interface, it will display all testing data of CV1.

                  進入“數據查詢”功能后,可通過上下移位按鍵對毛細管1當前歷史數據進行查詢,通過左右移位按鍵對毛細管1其它組歷史數據進行翻閱查詢,此時按下“打印”鍵可打印當前顯示的歷史數據。

                  After entered intodata querymenu, you can look up the current historical data of CV1 by shifting “Ù” “Ú” key, browse and inquiry of other data of CV1 by shifting×”“Økey, now click “print” will print all current displayed data.

                   

                  3.3試驗操作operations

                  在菜單界面下通過移位按鍵將光標移到“試驗操作”處,此時按“確定”鍵進入試驗操作界面。在試驗操作界面中,可對毛細管1~4進行測試。如下:

                  By shifting key move the cursor to “operations”, now click “enter” will link to operation interface. In this interface, it can test for 1~4capillary. As the following:

                  picture2-13

                  剛跳入此界面時,光標停在“毛細管1試驗”處,如果需要試驗,可直接按“確定”鍵,進入毛細管1試驗界面。如下:

                  The cursor stop at “CV1” when linked to this interface, if need to be tested, directly enter” key and link to corresponding testing interface. As the following:

                  picture2-4

                  在此界面下,如做好所有準備工作,當試樣液面回落到上刻度時按下“1”鍵開始計時,當試樣液面回落到下刻度時再次按下“1”鍵停止計時并鎖定時間值。如此反復測試四組試樣的流動時間后,自動計算出其平均流動時間和粘度值。測試結束后按下“打印”鍵可打印當前顯示的測試數據。

                  Under this interface, if ready for all, click1start to timing when the liquid level fall back to the upper scale, click1again to stop timing and locked the time when the liquid level fall back to the lower scale. Just like this repeatedly test the flow time of 4 teams automatically calculate the average flow time and viscosity value. After finishing the experiment, clickprintwill print the current displayed data.

                  在此過程中可穿插進行其它毛細管的試驗,操作方法相同,按對應的數字鍵起停,按][鍵可切換顯示不同毛細管的試驗數據。

                  During this process, you can alternately test other capillary with the same method, press corresponding number key control the start/stopclick[left][right] can switch to different displays of different capillaries.

                  3.4幫助help

                  在菜單界面下通過移位按鍵將光標移到“幫助”處,此時按“確定”鍵進入幫助界面。如下:

                  Under this interfaceby shifting key move the cursor to “help”, now click “enter” will link to help interface. As the following:

                  picture2-14

                      在幫助界面中,可查閱儀器型號、儀器名稱、軟件版本、公司名稱、公司網址。

                  In this interface, you can look up apparatus modelnamesoftware versioncompany name and website.

                  4 試驗步驟Procedures

                  4.1根據測定溫度的要求在浴缸內注入相應的浴液體至水位標記處。

                  According to measure temperature required, pour relevant bath liquid into the bath vat until it reached the marked level

                  4.2開機。Start up

                  4.3在“設置”菜單中根據樣品性能設置所需要的參數。

                  Under “setup” menu, according to samples performance, set the required parameters.

                  4.4安裝試樣:往內徑符合要求且清潔、干燥的毛細管粘度計內裝入試樣。在裝試樣之前,將橡皮管套在支管上,并用手指堵住管身的管口,同時倒置粘度計,然后將管身插入裝著試樣的容器中,這時利用橡皮球將液體吸到標線,同時注意不要使管身,擴張部分中的液體發生氣泡和裂隙。當液面達到標線時,就從容器里提起粘度計,并迅速恢復其正常狀態,同時將管身的管端外壁所沾著的多余樣擦去,并從支管取下橡皮管套在管身上。

                  Setup Sample: fill the samples into the cleandry suitable inner diameter capillary. Before setup the sample, cover the rubber casing on the branch tubes, jammed the nozzle with finger, meanwhile inverse the viscometer, then immerse the body into the container with samples, at this time sucked the liquids to the marked line with the rubber ball, beware of the liquids in dilated parts created air bubbles and gaps. When the level reached the marked line, pulls the viscometer out of the container, quickly restore its normal state, at the same time rub off the extra samples adhered to the external wall of tube end, and remove the rubber casing from the branch tubes and cover for the tube body.

                  4.5將粘度計管固定在粘度計管架上,并裝入水浴中。

                  Fixed the viscometer tube to the rack, and immerse it into the water bath.

                  4.6當實測溫度到達設定溫度后開始測試。利用毛細管粘度計管身口所套著的橡皮管將試樣吸入擴張部分,使試樣液面稍高于標線,并且注意不要讓毛細管和擴張部分的液體產生氣泡或裂隙。此時觀察試樣在管身中的流動情況,當液面正好回落到上標線時,按對應的“數字”鍵開始計時,液面正好回落到下標線時,再按對應的“數字”鍵停止計時,屏幕顯示本次試驗的運動時間。上次運動時間顯示后方可進行下一次的吸油,待第四次檢完后,屏幕除顯示四次運動時間外,儀器自動計算并顯示平均時間及最后的粘度結果。

                  Begin to test as the measured temperature reached the setting temperature. Make use of the rubber casing covered on the viscometer nozzle to draw the samples into extension parts, make the sample level more high than marked line, and prevent the liquids from air bubbles and gap. At this time, observe the flow condition of samples, when the level return to the upper marked line, click corresponding “number” key start to timing, when the level return to the lower marked line, click corresponding “number” key stop timing, the screen displayed the flow time of this test. The next extraction of oil can be conduct after displayed the last flow time, when completed four test, the apparatus automatically calculate and display average time and the last viscosity but displayed flow time of four tests.

                  4.7實驗結束后,對粘度計管進行清洗,并放在固定位置以便下一次試驗。

                  After experiment, clean up the viscometers and put it in a fixed location, so as to the next experiment.

                  5操作注意事項Attentions

                  5.1浴缸中的浴液不能有雜質與飄浮物。

                  There are no impurities and floating materials in bath liquid of the bath vat.

                  5.2注意要保持毛細管在水浴內垂直,可通過粘度計管架上的三個手擰螺絲進行微調。

                  Keep the capillary vertically the water bath; you can make a fine tuning through the 3screws in the rack.

                  5.4測定試樣的運動粘度時,應根據試驗的溫度選用適當的粘度計,務使試樣的流動時間不少于200s,內徑0.4mm的粘度計流動時間不少于350s

                  When testing the kinematic viscosity, you should select suitable viscometer according to the experimental temperature, make sure the flow time of samples is no less than 200s, and the flow time of viscometer with 0.4mm inner diameter is no less than 350s.

                  5.3在測定試樣的粘度之前必須將粘度計用溶劑油或石油醚洗滌,如果粘度計沾有污垢,就用鉻酸洗液、水、蒸餾水或95%乙醇依次洗滌。然后放入烘箱中烘干,或用通過棉花濾過的熱空氣吹干。

                  You must clean the viscometer with solvent oil or petroleum ether before testing the viscosity, if the viscometer has stains, Cleaning it with the chromate liquidwaterdistilled water or 95% Ethanol in turn. Then dry it in drier or hot air that was filtered with cotton.

                  5.5毛細管放入浴缸中后,必須保證在溫度到達規定值后15分鐘方可試驗,建議用戶在未到設定溫度前將毛細管放入浴缸中。

                  After put the capillary into the bath vat, you must guarantee didn’t conduct the experiment until the temperature has reach the stipulated value over 15minitues; it is suggest that put the capillary into the bath vat before reached the stipulated temperature.

                  5.6測試時油需吸到粘度計管上部球體的中間,并保證吸油粘度計管的毛細管常數與設定值相符。

                  When make the testing, oil need to be drawn into the middle of upper sphere, and to make sure that the coefficient of capillary is consistent with the setting value.

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                   

                  第三部分 儀器維護、保養及安全說明

                  Chapter Three  Maintenance & Safety Instructions

                  1儀器維護與保養Maintenance

                  1、  儀器應存放在干燥的地方,并作好防塵工作。

                  The apparatus should be store in dry conditionand make a good preparation for dust proof.

                  2、  開機前應確保儀器的各部件連接正常,請不要連續開關電源,以免損壞元器件,開關電源之間至少應間隔20S以上。

                  To ensure proper connection of apparatus before start upplease don’t switch the supply repeatedly in order to avoid damage electronic partsthe interval time should be at least 20s between the two switch.

                  3、  試驗前應做好安全防范工作,在儀器的周圍不應有可燃、易爆、腐蝕性物品。

                  Before conducting the experiment, you should make a good preparation for security, please don’t set combustibleexplosivecorrosive materials around the apparatus.

                  4、 玻璃缸先注水后方可通電使用。

                  The apparatus can be charged after the glass vat was filled into water.

                  5、面板為液晶顯示器,不能用硬物磕碰,否則會造成顯示器的永久性損壞。

                  The panel is LCD, please don’t scratch it with sharphard objects, Otherwise it will cause permanent damage.

                  6、儀器應放置在平整、穩固的機臺上并盡量避免陽光直射在儀器上。

                  The apparatus should be placed on a fixedstable platform and try to avoid straight sunshine.

                  7、如有誤操作,可按“復位“鍵,重新開始操作。

                  For any improper operations, click “reset” and restart.

                  8、電源必須做好良好接地。

                  The supply should be well grounded.

                  9試驗結束后應作好保潔工作,把粘度計管中的油樣清理干凈。

                  You should make a good job of cleaning after finishing the experiment, especially clean up the oil samples in the viscometer.

                  2粘度計管清洗Cleaning Viscometer

                  用鉻酸洗液、水、蒸餾水或95%乙醇對粘度計管依次洗滌。

                  Cleaning the viscometers with the chromate liquidwaterdistilled water or 95% Ethanol in turn.

                  3安全說明Safety Instructions

                  使用本機時,為了減少火災、觸電或人員傷害的危險,應始終遵守基本的安全預防措施。

                  In order to reduce the damage of fire hazardelectric shock or other injuries, you should abide by the following basic precautions when using the apparatus.

                  閱讀和理解全部說明。

                  Read and comprehended all instructions.

                  按照機身上注明的警告及說明進行操作。

                  Operate only under the guidance of the warning and instructions that marked on the apparatus.

                  在清潔本機前,請拔下本機的電源插頭。不要使用液體或噴霧清潔器。使用濕布進行清潔。   

                  Before cleaning, please remove the plug from the apparatus. Don’t use liquid or spray cleaner but wet cloth.                                    

                  將本機牢固地放置在穩定的地方。如果本機掉落,會導致嚴重損壞和/或人員傷害。

                  Put the apparatus in a fixed place. If it dropped, it might lead to serious damage or personnel injury.

                  請勿堵塞機器上的縫隙和開口。它們用于通風以防本機過熱。切勿將本機放置在散熱器附近或不能正常通風的地方。

                  Please don’t block the gap and breach in the apparatus, because it is used for ventilation so as to prevent the apparatus from hot. Never put it near radiator or other unventilated places.

                  本機只能使用機器上標明的電源。如果您不了解自己的供電類型,請向經銷商或當地供電局詢問。

                  The apparatus only can use the supply that marked in the unit. If you don’t know your supply type, please contact with local agent or power supply bureau.

                  為了安全,本機備有接地插頭。如果您沒有此種插座,請安裝一個。切勿不當修改插頭以免此安全功能不起作用。

                  The apparatus is equipped earth plug. If you don’t have this kind of socket, please install one. Don’t refit it so that the safety function lost.

                  為了減少觸電危險,切勿拆卸本機。當需要進行檢修時,請求技術支持。打開或取下蓋子可能會使您碰到危險的電壓或其他危險。不正確的重新組裝可能會在以后的使用中導致觸電。

                  In order to reduce the dangers of electric shock, please don’t dismantle the apparatus. If you need repair, please ask for technical support. You might get a shock or other dangerous when open or remove the cover. Incorrect reassemble will lead to electric shock in the process of using.

                  當發生下列情況時,請從電源插座拔出本機的插頭。

                  Under the following circumstance, please pull the plug from supply socket.

                    A.當電源線損壞或磨損時。

                  Power line has broken or worn.

                    B.如果有液體灑入本機。

                  There is liquid spill into the apparatus.

                    C.如果本機已遭受雨淋或水澆。

                  The apparatus has been drenched with rain or water.

                    D.如果按照使用說明書進行操作時本機不能正常工作。只能調節使用說明書中包括的控制部分。不正確的調節反而會使儀器故障擴大。

                  The apparatus can’t operate properly if you comply with instruction, you can only regulate the control parts. The problems will get worse due to improper regulation.

                    E.如果本機掉落或機身損壞。

                  The apparatus dropped or broke.

                    F.如果本機在性能上發生明顯的變化。

                  There is obvious change in performance.

                  注意:勿用濕手觸摸插頭。

                  Attention: Don’t touch the plug with a wet hand.